Hitachi Microscope & Magnifier S-3400N ユーザーズマニュアル
3.10.2
3 - 159
Table 3.10-3 Objective Lens Movable Aperture and Image Quality
Stop No.
1
2
3
4
Aperture diameter
(µm)
150 80
50 30
Focal depth
Shallow
Deep
Probe current
High
Low
Resolution Low
High
Application
X-ray analysis on
low accelerating
voltage
Observation and normal X-ray
analysis
Observations
requiring a great
focal depth
See 3.4.2 Axial Alignment.
3.10.4 Influence of WD in Low-vacuum Mode
When observation is conducted in the low-vacuum mode, in which the specimen chamber is
maintained at a low vacuum level, some of the primary electrons and the backscattered
electrons that provide signals are subject to scattering by residual gas molecules, and this may
result in a less than clear image. To ensure a clear image, the following settings may be required.
(1) Accelerating voltage
The accelerating voltage should be set as high as possible (see Table 3.10-1).
(2) Probe current
If image roughness is noticeable, the probe current should be increased (see Table 3.10-2).
(3) Working distance
To increase the sensitivity of the detector, the working distance should be minimized as
much as possible (see Table 3.10-4).
much as possible (see Table 3.10-4).
Table 3.10-4 Impact of WD in Low-vacuum Mode
Working Distance (mm)
10
50
Scattering of irradiation electrons
Low
High
Scattering of backscattered electrons
Low
High