6.3.2 
6 - 11 
6.3.2  Each Command 
 
Table 6.3-1 Property Command 
 
Command Return  Contents 
IsActive 
0 : SEM not execute 
1 : Ready 
SEM status reading 
HVOn 
0 : HV-OFF 
1 : HV-ON 
HV ON/OFF monitoring 
Vacc 
0 : HV
  OFF 
300-30000 : 0.3-30 kV 
Acceleration voltage 
monitoring 
Mag 5-300000
:x5-x300k Magnification. 
monitoring 
WD 400-7000
:4.00-70.00 mm  W.D. monitoring 
ObjLensCoarse 
0-16383 
Focus course value 
monitoring 
ObjLensFine 
0-4095 
Focus fine value monitoring
ProbeCurrent 
0-1000 : 0.0-100.0 
Probe current monitoring 
ChamberType 
0 : Standard specimen 
chamber 
Chamber type reading 
StageType 
0 : Manual stage 
1 : 5 axis motor driven 
stage 
Stage type reading 
VacuumMode 
0 : High vacuum 
1 : Low vacuum 
Vacuum mode reading 
Vacuum 
6-270 : 6-270Pa 
Vacuum value monitoring 
InstructName 
“S-3400N” 
Instrument name reading 
Version 
0-20 
Software version reading 
Revision 
0-99 
Software revision reading