Code 
Message Text 
Explanation and Countermeasure 
7320  An abnormal power supply 
(Vacc) in the electron gun control 
power supply was detected.   
Shut down the PC and turn the 
[MAIN] switch off.   
Restart the microscope after 
about half an hour. 
When this message is shown 
again, contact a service engineer.
7321  An abnormal power supply 
(Vfilament) in the electron gun 
control power supply was 
detected.  
Shut down the PC and turn the 
[MAIN] switch off.   
Restart the microscope after 
about half an hour. 
When this message is shown 
again, contact a service engineer.
7322  An abnormal power supply 
(Vbias) in the electron gun control
power supply was detected.   
Shut down the PC and turn the 
[MAIN] switch off.   
Restart the microscope after 
about half an hour. 
When this message is shown 
again, contact a service engineer.
7323  An abnormal power supply 
(Vbiasselect) in the electron gun 
control power supply was 
detected.  
Shut down the PC and turn the 
[MAIN] switch off.   
Restart the microscope after 
about half an hour. 
When this message is shown 
again, contact a service engineer.
This message appears if the electron gun-controlling power 
supply unit detects an error during a self test.    It is possible 
that the error is due to malfunction or equipment error from 
isolated noises, in which case shut down the PC, turn off 
the Main switch, wait 30 minutes, and then restart the 
system. If the problem persists, call a service engineer. 
7324  The electron gun interlock switch 
worked.  
Shut down the PC, turn the 
[MAIN] switch off and contact a 
service engineer. 
This message appears if the interlock switch for the 
electron gun unit is activated when the acceleration voltage 
is still on. Shut down the PC, turn off the Main switch, and 
then call a service engineer. 
7325  Over current protection of the 
electron gun operated. Please 
confirm the setting of the 
filament. 
This message appears if the protection circuit is activated 
due to a surge in emission current due to a filament height 
settings. See the filament replacement/servicing 
procedures in the Help menu to check the filament settings.