2.3.6 
2 - 35 
2.3.6.8  EVAC/AIR Control 
 
 
 
Fig. 2.3-21  EVAC/AIR Control 
 
Clicking the EVAC button initiates the evacuation of the specimen chamber.    Similarly, clicking 
the AIR button causes the AIR processing of the specimen chamber. 
The display window to the left of the EVAC button lights up when the specimen chamber is in the 
EVAC state; it blinks when the EVAC process is in progress; and in other cases it remains off.   
Similarly, the display window to the left of the AIR button lights up when the specimen chamber 
is in the AIR state; it blinks when the AIR processing is in progress; and it remains off in all other 
cases. 
 
Status 
EVAC 
Button 
EVAC 
Window 
AIR 
Button 
AIR 
Window 
EVAC fall sequence 
Disabled 
※1 Disabled ※1 
EVAC rise sequence 
Disabled 
※2 Disabled ※2 
EVAC processing on specimen 
chamber  
Disabled Blinking  Enabled 
Off 
Specimen chamber in HVON/OFF 
operation in EVAC mode 
Disabled Lit Disabled Off 
Specimen chamber undergoing an 
EVAC mode change in EVAC mode 
Disabled Blinking  Enabled 
Off 
Specimen chamber in EVAC mode, 
and in LOW-EVAC mode the 
vacuum level currently set does not 
agree with the current vacuum level. 
Disabled Blinking 
※3 
Enabled Off 
Specimen chamber in EVAC mode 
and in conditions other than the 
above 3 items 
Disabled Lit  Enabled Off 
Specimen chamber undergoing AIR 
processing  
Enabled Off Disabled 
Blinking 
Specimen chamber in AIR   
Enabled 
Off 
Disabled 
Lit 
 
※1:  Both EVAC and AIR windows blink alternately. 
※2:  The applicable window blinks, depending on the EVAC/AIR condition. 
※3:  Only the EVAC window blinks in deep green.