Hach Portable Spectrophotometer DR 2800 ユーザーズマニュアル

ページ / 126
37
Standard Operations
5.3 Store, recall, send and delete data 
5.3.1 The data log
The Data Log will store up to 500 readings taken in the modes: 
Stored Programs, Barcode Programs, User Programs, Favorite 
Programs, Single Wavelength and Multi Wavelength. A complete 
record of the analysis is stored, including the Date, Time, Results, 
Sample ID, and Operator ID.
5.3.1.1 Auto/manual data storage
The data storage parameter indicates whether data are to be stored 
automatically or manually (in which case the user has to decide 
which data to store).
1. Press Store: On/Off in the Options menu.
With the Store On setting, all measurement data are stored 
automatically. 
With the Store Off setting, no measurement data are 
stored. However, this setting can be changed to Store On 
in the result display through configuration. The reading 
currently shown in the display is then stored.
Note: When the instrument's memory (data log) is full, the oldest data are 
automatically deleted allowing the new data to be stored.
5.3.1.2 Recall stored data from the data log
1. Press Recall Data in the Main Menu.
2. Press Data Log.
A listing of the stored data is displayed.
3. Press Filter: On/Off