Hach 900 MAX ユーザーズマニュアル

ページ / 160
Page 24
Installing the Pump Tube in the Sensor Body
8990hrd.fm
Section 2
2.3 Installing the Pump Tube in the Sensor Body
Note: Do not stretch the tubing in 
the sensor body, as this could affect 
the ability of the sensor to detect 
liquid through the pump tubing.
1.
Remove the four screws on the pump cover (
). 
2.
Remove the front cover of the pump housing. Remove the tubing. Locate 
the black dots on the tubing. The end of the tube that extends farthest 
beyond the black dot attaches to the stainless steel tubing connector. 
3.
Install the pump tube in the pump housing so the black dots are visible 
just outside the pump body.
Note: Use the proper length of 
silicone tubing in the pump body. An 
improper length can reduce the life 
of the tubing and pump rollers. Refer 
to 
 for the correct length.
4.
After inserting the new pump tube as shown, reinstall the front cover and 
secure it with the four screws until finger tight.
5.
Make sure that the tubing extends through the liquid sensor and out of the 
controller as shown in 
.
Figure 5
Pump Tube Loading 
Figure 6
Installing Pump Tube Through the Sensor Body
To Intake Tubing Connector
5 3/4 in. to Sample Fitting
11 5/8 in.
(Tubing in Pump)