Hach 900 MAX ユーザーズマニュアル

ページ / 160
Section 3
Page 39
8990startup.fm
Basic Programming Setup
1-C. Enter the bottle volume using the numeric keypad and select gallons or 
milliliters using the 
CHANGE UNITS 
key.
1-D. Press 
ACCEPT
 and continue to Intake Tubing.
Step 2 - Intake Tubing
2-A. Enter the intake tube length attached to the sampler, using the numeric 
keypad. Length values from 100 to 3000 cm (3 to 99 ft) are valid. 
Change the measurement unit using the 
CHANGE UNITS
 key.
Note: Intake tube length affects sample volume accuracy, cut the tubing to the nearest 
whole foot. Avoid excessively long tubing runs. Doubling the intake tubing length 
can quadruple the pump tubing wear if intake rinses or sample retries are 
enabled. Always locate the sampler close to the sample liquid source to minimize 
tubing, pump, motor, and gearbox maintenance.
2-B. Press 
ACCEPT
 to move to the Intake Tube Type menu.
2-C. Select the type of intake tube (
3
/
8
 in. Vinyl, ¼ in. Vinyl, 
3
/
8
 in. Teflon
®
using the 
CHANGE UNITS 
key.
2-D. Press 
ACCEPT
 to select the intake tube type and continue with Program 
Lock.
Step 3 - Program Lock
Note: The program lock password is 
configured at the factory as “9000” 
and cannot be changed.
Enabling the Program Lock will provide a protective “password” which will 
keep unauthorized personnel from tampering with the instrument keypad.
3-A. Enable or Disable Program Lock using the 
CHANGE CHOICE 
key. 
Note: When program lock is enabled and a user attempts to make a change in the 
program, a screen will ask the operator to enter the password. The operator must 
enter 9000 and press ACCEPT.
3-B. Press 
ACCEPT
 to continue the basic program setup and continue with 
Program Delay.
11:00 AM 21 - APR - 01 
BOTTLES
ACCEPT
BOTTLE VOLUME
CHANGE
UNITS
— —
BACKUP
CLEAR
ENTRY
ENTER: 0.500—99.90
11:00 AM 21 - APR - 01 
INTAKE TUBING
ACCEPT
INTAKE TUBE LENGTH:
CHANGE
UNITS
_____
BACKUP
CLEAR
ENTRY
ENTER: 3—99