Step 12 - Program Complete Output
Program Complete Output sends a +12 V dc signal out Pin F of the Auxiliary
Receptacle at the completion of the sampling program. This signal is also sent
when a Full Bottle condition causes the program to complete. The Program
Complete signal remains on (+12 V dc) for 61 seconds and then turns
back off (0 V dc).
Receptacle at the completion of the sampling program. This signal is also sent
when a Full Bottle condition causes the program to complete. The Program
Complete signal remains on (+12 V dc) for 61 seconds and then turns
back off (0 V dc).
Program Complete Output is used for the following purposes:
•
Multiple Sampler Operation. Samplers are arranged in a “Cascade”
control where the first sampler signals a second sampler to start its
program when the first sampler’s program is complete. This feature
requires the second sampler to be programmed for external
start operation.
control where the first sampler signals a second sampler to start its
program when the first sampler’s program is complete. This feature
requires the second sampler to be programmed for external
start operation.
•
To interface to a PC to signal the end of the sampling program.
•
To interface to a relay that trips a signal light in a control room to indicate
that sampling is finished and samples are ready to be picked up.
that sampling is finished and samples are ready to be picked up.
12-A. From the Advanced Sampling menu highlight Program Complete using
the
UP
and
DOWN
keys. Press
SELECT
to continue.
12-B. Enable or Disable Program Complete using the
CHANGE CHOICE
key.
12-C. Press
ACCEPT
to continue to Set Point Sampling.
Step 13 - Setpoint Sampling
Setpoint sampling allows the control of an automatic liquid sampler from one
of up to 14 sources. Setpoint sampling defines a set of limits that inhibit
sampling until an upset condition occurs, causing the limits to exceed.
Sampling is enabled only when the waste stream falls outside the setpoints.
of up to 14 sources. Setpoint sampling defines a set of limits that inhibit
sampling until an upset condition occurs, causing the limits to exceed.
Sampling is enabled only when the waste stream falls outside the setpoints.
13-A. Highlight Setpoint Sampling using the
UP
and
DOWN
keys on the
Advanced Sampling Menu. Press
SELECT
.
13-B. Enable or Disable Setpoint Sampling using the
CHANGE CHOICE
key.
Press
ACCEPT
to continue.
11:00 AM 21 - APR - 01
ADVANCED SAMPLING
SELECT
PGM CMPLT OUTPUT
RETURN
SETPOINT SAMPLING
SPECIAL OUTPUTS
SPECIAL OUTPUTS
11:00 AM 21 - APR - 01
ADVANCED SAMPLING
SELECT
PGM CMPLT OUTPUT
SETPOINT SAMPLING
RETURN
SPECIAL OUTPUTS
START/STOP TIMES
START/STOP TIMES