Hach 900 MAX ユーザーズマニュアル

ページ / 160
Section 4
Page 63
8990sensors.fm
Submerged Area/Velocity Sensor
4.
Select either inches or centimeters using the
 CHANGE UNITS
 key. The 
distance must be greater than the minimum deadband of 10 in. (25.4 cm) 
for the 75 kHz sensor and 15 in. (38.1 cm) for the 50 kHz sensor.
5.
Press 
ACCEPT
 when finished. 
4.2 Submerged Area/Velocity Sensor
The Submerged area/velocity sensor simultaneously measures level 
and velocity.
4.2.1 Submerged Area/Velocity Sensor Connection
Recommended Routing of Submerged Area/Velocity Cable
Proper routing of the cable serves two functions: 
Keeps the in-line desiccant assembly inside the cover and out of the way 
when installing and removing the sampler from tight manholes.
Keeps the desiccant out of direct contact with the elements. Also, keeps 
water from dripping directly onto the desiccant assembly.
To secure the submerged area velocity cable, route the cable around the 
battery and clip it under the battery hold downs.
4.2.2 Submerged Area/Velocity Sensor Programming
1.
From the Main Menu, select 
OPTIONS>LEVEL SENSOR.
2.
Select Submerged Xducer using the 
CHANGE CHOICE
 key. Press 
ACCEPT
.
3.
From the 
MAIN MENU
, select 
SETUP>MODIFY SELECTED ITEMS.
4.
Highlight Velocity Direction using the 
UP
 and 
DOWN 
keys. Press 
SELECT
.
5.
Set the velocity direction (upstream, downstream, or always positive) 
using the 
CHANGE CHOICE
 key. Press 
ACCEPT
 to continue.
6.
Highlight Velocity Units using the 
UP
 and 
DOWN 
keys. Press 
SELECT
.
Table 5 Submerged/Area Velocity Sensor Connection
Pin
Signal Description
Wire Color
Bare Leads
A
+12 V dc
Red
Red
B
ground
Green
Green
C
Receive (ground)
B/W Shield
Yellow
D
Receive (+)
B/W Center
Orange
E
Transmit (ground)
Black Shield
Violet
F
Transmit (+)
Black Center
Gray
G
Depth (-)
Black
Black
H
Depth (+)
White
White