Hach 900 MAX ユーザーズマニュアル

ページ / 160
Page 90
Alarm Relays
8990cm.fm
Section 6
6.4.1 Alarm Relays Connection
Figure 26
Relay Pin Connections
Relay Junction Box
Required relay box is an alarm relay box assembly with a 10-ft long cable with 
a 6-pin connector on one end and a relay box on the other end (
1.
Choose Normally Closed (NC) or Normally Open (NO) connections. 
2.
Connect one wire to the common (COM) and the other to the connector 
with the preferred signal.
Table 17 Relay Connector (J17)
Pin
Signal Description
Wire Color
A
+12  V  dc
Red
B
Relay #1
Yellow
C
Relay #2
Black
D
Relay #3
Red
E
Relay #4
Green 
Table 18 Relays
Connector
Relay
J2
1
J3
2
J4
3
J5
4