Troubleshooting
Proline Promass 80 PROFIBUS PA
78
702
P: MEDIUM INHOM.
!: # 702
0x43
UNCERT
AIN
Non specific
(uncertain 
status)
Constant
Fluid 
inhomogeneous
Cause:
Frequency control is not stable, due to inhomogeneous 
process fluid, e.g. gas or solid content.
Remedy:
In particular with outgassing fluids and/or increased 
gas content, the following measures are recommended 
to increase system pressure:
– Install the instrument at the outlet side of a pump.
– Install the instrument at the lowest point of an 
ascending pipeline.
– Install a flow restriction, e.g. reducer or orifice plate, 
downstream from the instrument.
703
P: NOISE LIM. CH0
!: # 703
0x40
0x41
0x42
UNCERT
AIN
Non specific
(uncertain 
status)
O.K.
Low
High
Noise Limit 
Channel 0
Cause:
Overdriving of the internal analog to digital converter.
Causes:
Cavitation
Extreme pressure pulses
High gas flow velocity
A continuation of the measurement is still possible!
Remedy:
Change or improve process conditions, e.g. by 
reducing the flow velocity.
704
P: NOISE LIM. CH1
!: # 704
0x40
0x41
0x42
UNCERT
AIN
Non specific
(uncertain 
status)
O.K.
Low
High
Noise Limit 
Channel 1
705
P: FLOW LIMIT
$: # 705
0x42
UNCERT
AIN
Non specific
(uncertain 
status)
High 
Limited
Flow limit
Cause:
The mass flow is too high. The electronics' measuring 
range will be exceeded.
Remedy:
Reduce flow
731
P: ADJ. ZERO FAIL
!: # 731
0x43
UNCERT
AIN
Non specific
(uncertain 
status)
Constant
Zero point 
adjustment failure
Cause:
The zero point adjustment is not possible or has been 
canceled.
Remedy:
Make sure that zero point adjustment is carried out at 
"zero flow" only (v = 0 m/s). 
No.
Device status message
(local display)
PROFIBUS measured value status
Extended 
diagnostic 
message in the 
PROFIBUS 
master
Cause/remedy
Q
u
a
lit
y co
de (H
EX
)
M
e
asured va
lue st
atus
Qua
li
ty st
a
tus
Qu
ality
 substatus
Li
m
its