Agilent Technologies Printer Accessories 1670G ユーザーズマニュアル
![Agilent Technologies](https://files.manualsbrain.com/attachments/f20ae6efbee7636e6d0d1a481fd9e3847e3caeb5/common/fit/150/50/991587f279eee8cef3aa0a068e5c2f53d86b00e1bf17fbcbf9a87aa49d96/brand_logo.png)
Query
:MACHine{1|2}:SLISt:MMODe?
The MMODe query returns the current marker mode selected.
Returned Format
[:MACHine{1|2}:SLISt:MMODe] <marker_mode><NL>
Example
OUTPUT XXX;":MACHINE1:SLIST:MMODE?"
OPATtern
Command
:MACHine{1|2}:SLISt:OPATtern
<label_name>,<label_pattern>
<label_name>,<label_pattern>
The OPATtern command allows you to construct a pattern recognizer term
for the O Marker which is then used with the OSEarch criteria when moving
the marker on patterns. Because this command deals with only one label at a
time, a complete specification could require several invocations.
for the O Marker which is then used with the OSEarch criteria when moving
the marker on patterns. Because this command deals with only one label at a
time, a complete specification could require several invocations.
When the value of a pattern is expressed in binary, it represents the bit
values for the label inside the pattern recognizer term. In whatever base is
used, the value must be between 0 and 2
values for the label inside the pattern recognizer term. In whatever base is
used, the value must be between 0 and 2
32
−
1, since a label may not have
more than 32 bits. Because the
<label_pattern>
parameter may contain
don’t cares, it is handled as a string of characters rather than a number.
<label_name>
string of up to 6 alphanumeric characters
<label_pattern>
"{#B{0|1|X} . . . |
#Q{0|1|2|3|4|5|6|7|X} . . . |
#H{0|1|2|3|4|5|6|7|8|9|A|B|C|D|E|F|X} . . . |
{0|1|2|3|4|5|6|7|8|9} . . . }"
Example
OUTPUT XXX;":MACHINE1:SLIST:OPATTERN ’DATA’,’255’ "
OUTPUT XXX;":MACHINE1:SLIST:OPATTERN ’ABC’,’#BXXXX1101’ "
SLISt Subsystem
OPATtern
17–11