the User Defined option when setting the trigger position from the front
panel.
panel.
The TPOSition command is only available when the acquisition mode is set to
manual.
manual.
<time_val>
real number from (2
×
sample_period) to (516096
×
sample_period)
<percent>
integer from 1 to 100
Example
OUTPUT XXX;":MACHINE2:TWAVEFORM:TPOSITION CENTER"
Query
MACHine{1|2}:TWAVeform:TPOSition?
The TPOSition query returns the current trigger setting.
Returned Format
[MACHine{1|2}:TWAVeform:TPOSition] {STARt|CENTer|END|DELay,
<time_val>| POSTstore,<percent>}<NL>
Example
OUTPUT XXX;":MACHINE2:TWAVEFORM:TPOSition?"
VRUNs
Query
:MACHine{1|2}:TWAVeform:VRUNs?
The VRUNs query returns the number of valid runs and total number of runs
made. Valid runs are those where the pattern search for both the X and O
markers was successful resulting in valid delta time measurements.
made. Valid runs are those where the pattern search for both the X and O
markers was successful resulting in valid delta time measurements.
Returned Format
[:MACHine{1|2}:TWAVeform:VRUNs] <valid_runs>,<total_runs><NL>
<valid_runs>
zero or positive integer
<total_runs>
zero or positive integer
Example
OUTPUT XXX;":MACHINE1:TWAVEFORM:VRUNS?"
TWAVeform Subsystem
VRUNs
VRUNs
23–20