MODE
Command
:SPA{1|2}:MODE {OVERView|HISTogram|TINTerval}
The MODE command selects which menu to display: State Overview, State
Histogram, or Time Interval. A query returns the current menu mode.
Histogram, or Time Interval. A query returns the current menu mode.
Example
OUTPUT XXX;":SPA1:MODE OVERView"
OUTPUT XXX;":SPA2:MODE HISTogram"
OUTPUT XXX;":SPA1:MODE TINTerval"
Query
:SPA{1|2}:MODE?
Returned Format
[:SPA{1|2}:MODE] {OVERView|HISTogram|TINTerval}<NL>
Example
10 DIM String$[100]
20 OUTPUT XXX;":SELECT 1"
30 OUTPUT XXX;":SPA1:MODE?"
40 ENTER XXX;String$
50 PRINT String$
60 END
SPA Subsystem
MODE