TINTerval:TINTerval
Command 
:SPA{1|2}:TINTerval:TINTerval
<interval_number>,<min_time>,<max_time>
The TINTerval:TINTerval command specifies the minimum and maximum
time limits for the given interval.  A query returns these limits for a specified
interval.
<interval_
number>
0 to 7
<min_time>
real number
<max_time>
real number
Example
OUTPUT XXX;":SPA2:TINTerval:TINTerval 4,1.0E-3,47.0E5"
OUTPUT XXX;":SPA1:TINTerval:TINTerval 3,6.8E-7,4.90E2"
Query 
:SPA{1|2}:TINTerval:TINTerval? <interval_number>
Returned Format
[:SPA{1|2}:TINTerval:TINTerval ]<interval_number>,<min_time>,
<max_time><NL>
Example
10 DIM String$[100]
20 OUTPUT XXX;":SELECT 1"
30 OUTPUT XXX;":SPA2:TINTerval:TINTerval? 6"
40 ENTER XXX;String$
50 PRINT String$
60 END