OFFSet 
Command 
:CHANnel<N>:OFFSet <value>
The OFFSet command  sets the voltage that is represented at center screen
for the selected channel.  The allowable offset voltage values are shown in
the table below.  The table represents values for a Probe setting of 1:1.  The
offset value is recompensated whenever the probe attenuation factor is
changed.
<N> {1|2}
<value>
allowable offset voltage value shown in the table below.
Vertical Range
Offset Voltage
16 mV - 400 mV 
±
2 V
>400 mV - 1.6 V 
±
10 V
>1.6 V - 10 V 
±
50 V
>10 V - 40 V
±
250 V
Example
OUTPUT XXX;":CHAN1:OFFS 1.5"
Query 
:CHANnel<N>:OFFSet? 
The OFFSet query returns the current value for the selected channel.
 Returned Format
[:CHANnel<N>:OFFSet] <value><NL>
Example
OUTPUT XXX;":CHANNEL1:OFFSET?"
CHANnel Subsystem
OFFSet 
30-6