NWIDth?
Query
:MEASure:[SOURce CHANnel<N>;]NWIDth?
The NWIDth query makes a negative width time measurement on the
selected channel. The measurement is made between the 50% points of the
first falling and the next rising edge displayed on screen. If a parameter
cannot be measured, the instrument responds with 9.9E37.
selected channel. The measurement is made between the 50% points of the
first falling and the next rising edge displayed on screen. If a parameter
cannot be measured, the instrument responds with 9.9E37.
Returned Format
[:MEASure:NWIDth] <value><NL>
<N>
{1|2}
<value>
negative pulse width in seconds
Example
OUTPUT XXX;":MEASURE:SOURCE CHAN2;NWID?"
OVERshoot?
Query
:MEASure:[SOURce CHANnel<N>;]OVERshoot?
The OVERshoot query makes an overshoot measurement on the selected
channel. The measurement is made by finding a distortion following the first
major transition. The result is the ratio of OVERshoot to VAMPlitude. If
either cannot be measured, the instrument responds with 9.9E37.
channel. The measurement is made by finding a distortion following the first
major transition. The result is the ratio of OVERshoot to VAMPlitude. If
either cannot be measured, the instrument responds with 9.9E37.
Returned Format
[:MEASure:OVERshoot]<value><NL>
<N>
{1|2}
<value>
ratio of OVERshoot to VAMPlitude
Example
OUTPUT XXX;":MEASURE:SOURCE CHAN1;OVER?"