NWIDth?
Query
:MEASure:[SOURce CHANnel<N>;]NWIDth?
The NWIDth query makes a negative width time measurement on the
selected channel.  The measurement is made between the 50% points of the
first falling and the next rising edge displayed on screen.  If a parameter
cannot be measured, the instrument responds with 9.9E37.
Returned Format
[:MEASure:NWIDth] <value><NL>
<N>
{1|2}
<value>
negative pulse width in seconds
Example
OUTPUT XXX;":MEASURE:SOURCE CHAN2;NWID?"
OVERshoot?
Query
:MEASure:[SOURce CHANnel<N>;]OVERshoot?
The OVERshoot query makes an overshoot measurement on the selected
channel.  The measurement is made by finding a distortion following the first
major transition.  The result is the ratio of OVERshoot to VAMPlitude.  If
either cannot be measured, the instrument responds with 9.9E37.
Returned Format
[:MEASure:OVERshoot]<value><NL>
<N>
{1|2}
<value>
ratio of OVERshoot to VAMPlitude
Example
OUTPUT XXX;":MEASURE:SOURCE CHAN1;OVER?"