Query
:MEASure:[SOURce CHANnel<N>;]VMAX?
The VMAX query returns the absolute maximum voltage of the selected
source. If a parameter cannot be measured, the instrument responds with
9.9E37.
source. If a parameter cannot be measured, the instrument responds with
9.9E37.
Returned Format
[:MEASure:VMAX] <value><NL>
<N>
{1|2}
<value>
maximum voltage of selected waveform
Example
OUTPUT XXX;":MEASURE:SOURCE CHAN2;VMAX?"
VMIN?
Query
:MEASure:[SOURce CHANnel<N>;]VMIN?
The VMIN query returns the absolute minimum voltage present on the
selected source. If a parameter cannot be measured, the instrument
responds with 9.9E37.
selected source. If a parameter cannot be measured, the instrument
responds with 9.9E37.
Returned Format
[:MEASure VMIN] <value><NL>
<N>
{1|2}
<value>
minimum voltage of selected waveform
Example
OUTPUT XXX;":MEASURE:SOURCE CHAN1;VMIN?"
MEASure Subsystem
VMAX?
33-11