Agilent Technologies Printer Accessories 1670G ユーザーズマニュアル

ページ / 626
VALid?
Query
:WAVeform:VALid?
The VALid query checks the oscilloscope for acquired data.  If a
measurement is completed, and data has been acquired by all channels, then
the query reports a 1.  A 0 is reported if no data has been acquired for the
last acquisition.
Returned Format
[:WAVeform:VALid] {0|1}<NL>
0
No data acquired
1
Data has been acquired
Example
OUTPUT XXX;":WAVEFORM:VALID?"
XINCrement?
Query
:WAVeform:XINCrement?
The XINCrement query returns the X increment currently in the preamble.
This value is the time difference between the consecutive data points.
X increment is determined by the RECord mode as follows:
In FULL record mode, the X-increment equals the time period between
data samples (or sample period).
In WINDow record mode, the X increment is the time between data points
on the display.  The X increment for WINDow record data will be less than
or equal to the sample period.
Returned Format
[:WAVeform:XINCrement]<value><NL>
<value>
X increment value currently in preamble
Example
OUTPUT XXX;":WAVEFORM:XINCREMENT?"
WAVeform Subsystem
VALid?
36-13