A&D MF-50 ユーザーズマニュアル

ページ / 75
MS-70 / MX-50 / MF-50 / ML-50 
35 
Select an Analysis Mode
Select an Analysis Mode
Select an Analysis Mode
Select an Analysis Mode    
 5. Press 
the 
 SELECT 
  key to select a symbol, it 
will blink. Select 
 U-t 
  of the timer mode with 
the 
 or 
 key. (Press these keys to select it) 
 
 
Set the Heating Pattern
Set the Heating Pattern
Set the Heating Pattern
Set the Heating Pattern    
  6.  Select drying temperature with the 
 SELECT 
 key. 
Select ramp drying 
  of the heating pattern 
with the 
 PROGRAM 
 key. 
 
   
 
 
Set the Final Drying Temperature
Set the Final Drying Temperature
Set the Final Drying Temperature
Set the Final Drying Temperature    
 
7.  Set 160°C with the 
 or 
 key. 
 
 
 
Set the Ramp Time
Set the Ramp Time
Set the Ramp Time
Set the Ramp Time    
 
8.  Set 5.0 minutes with the 
 or 
 key. 
 
 
 
Set the Analysis Time
Set the Analysis Time
Set the Analysis Time
Set the Analysis Time    
 
9.  Set 10.0 minutes with the 
 or 
 key. 
 
 
 
Set a Measurement Unit
Set a Measurement Unit
Set a Measurement Unit
Set a Measurement Unit    
  10.  Select a measurement unit with the 
 SELECT 
 
key. Select moisture content (based on a wet 
sample ) with the 
 or 
 key. 
 
 
Set the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % Display    
  11.  Select the % display with the 
 SELECT 
 key. 
Select 0.01 [%] with the 
 or 
 key. 
 
 
Set the Minimum Scale Value of the G
Set the Minimum Scale Value of the G
Set the Minimum Scale Value of the G
Set the Minimum Scale Value of the Gram Display
ram Display
ram Display
ram Display    
  12.  Select gram display with the 
 SELECT 
 key. 
Select 0.001 [g] with the 
 or 
 key. 
 
 
Store the Parameters and Finish the Operation
Store the Parameters and Finish the Operation
Store the Parameters and Finish the Operation
Store the Parameters and Finish the Operation    
 13.  Press the 
 ENTER 
  key to store the new parameters 
of the measurement program to program number 4. 
Pressing the key, the weighing mode is 
automatically displayed. When PROG 4 is recalled, 
the settings can be used. 
 
  To cancel the new parameters and return to the 
weighing mode, press the 
 RESET 
 key. 
"U"-"t
 0.000
g
 
Analysis mode 
    
 0.000
g
 
"1"0"5°C 
Ramp Time 
 
  "5."0 min 
    
 
   
0
"% 
"M"O"I"S"T 
    "/"W 
   
 
Measurement unit 
  
"0."0"1
%
 
   
Minimum scale 
value of % display 
 "0."0"0"1
g
 
   
Minimum scale 
value of gram 
display 
 0.000
g
 
160°C 
10.0 min 
 
"1"0."0 min 
Analysis time 
 0.000
g
 
"1"6"0°C 
Final drying temperature 
Ramp drying 
Gram display of 