Bosch Appliances Home Care Product F.01U.252.722 ユーザーズマニュアル

ページ / 36
22
en | Programming
NurseCall Relay Unit
F.01U.252.722 | V1 | 2011.12
User Manual
Bosch Security Systems
4.3.4
Assistance and fire non priority
This command sets the assistance call and the fire alarm as non-priority calls. This means that 
the last alarm is displayed, whatever its type. This is the default value.
1.
Type the code 123992. The unit plays a melody and a confirmation message is displayed.
4.3.5
Special texts in German
This command sets special texts in German. The displayed criteria are:
 BAD/WC instead of TECHNIK
HILFE-2 instead of NOTRUF2
1.
Type the code 123007. The unit plays a melody and a confirmation message is displayed.
4.3.6
Standard texts in German
This command sets standard texts in German. This is the default value.
1.
Type the code 123008. The unit plays a melody and a confirmation message is displayed.
4.4
Printer/display setup on the Relay Plus Unit
4.4.1
Access to the printer/display setup menu
1.
Press 
 then 
 to access the programming menu.
2.
Press 
 to enter the RS-232 output menu or 
 to check the RS-232 output setup.
3.
Press 
 to enter the printer/display setup. Press 
 to escape.
4.
Scroll with 
 and 
 to select one of the following:
None
Printer
DISPLAY ZETTLER
DISPLAY NORSK
ASSISTANCE & FIRE
NONPRIORITY
MULTITONE TEXTE
BAD/WC + HILFE-2
STANDARDTEXTE
TECHNIK + NOTRUF2
NOTICE! 
The printer/display setup can only be configured on a NurseCall Relay Plus Unit connected to 
a NurseCall system via the RS-485 bus.
¬
OK:Program
¤:Info
¬ Parameter Nr. 00
RS-232 Output
£