Omron Healthcare COMPAIR NE-C28 ユーザーズマニュアル

ページ / 112
10
Operating Instructions
PREPARING THE NEBULIZER FOR USE
5. Lift up the baffle and remove the baffle from the medication cup.
CAUTION
Do not alter the baffle, the nozzle in the medication cup or 
any part of the nebulizer kit.
6. Add the correct amount of prescribed medication to the 
medication cup.
WARNING
For type, dose, and regime of medication follow the 
instructions of your physician or licensed healthcare practitioner.
WARNING
Do not spill water or other liquids on the compressor and 
power cord. The compressor and power cord are not 
waterproof. If liquid spills on these parts, immediately 
unplug the power cord and wipe off the liquid with gauze or other 
soft absorbent material.
WARNING
Do not use tap or mineral water in the nebulizer for 
nebulizing purposes.
CAUTION
Do not add more than 7ml of medication to the medication cup.
7. Place the baffle back into the medication cup. Make sure 
the baffle is correctly positioned over the nozzle in the 
medication cup.
8. Place the inhalation top on the medication cup over the 
baffle. The top of the baffle should be centered in the 
inhalation top.
Align the protrusion on the inhalation top with the 
indentation on the medication cup as illustrated (A). 
Turn the inhalation top clockwise until securely closed (B).  
max
5
3