SCXI-1125 Calibration Procedure
calOp
ND_SELF_CALIBRATE
setOfCalConstants
ND_USER_EEPROM_AREA
calRefVolts—0.0
4.
Call
AI_Configure
to put the E Series device in bipolar mode. Set the
following parameters:
deviceNumber—The value assigned by MAX
chan—0
inputMode—0
inputRange—0
polarity—0
driveAIS—0
5.
Call
MIO_Config
to enable dithering on the E Series device. Set the
following parameters:
deviceNumber—The value assigned by MAX
dither—1
useAMUX—0
6.
Call
SCXI_Single_Chan_Setup
to configure the module for
single-channel measurements. Set the following parameters:
SCXIchassisID—The value assigned by MAX
moduleSlot—1
moduleChan—The channel you want to adjust
DAQdeviceNumber—The device number assigned by MAX for
the E Series DAQ device
7.
Call
SCXI_Set_Gain
.
 
Set the following parameters:
SCXIchassisID—The ID of the SCXI chassis obtained when you
configure the chassis with MAX
moduleSlot—1, because the SCXI-1125 should be installed in
slot 1
channel—The same channel number from step 6
gain—The gain value you want to adjust. Refer to Table 1 for
acceptable gain values
8.
Call
SCXI_Configure_Filter
to configure the filter setting on the
module. Set the following parameters:
SCXIchassisID—The ID of the SCXI chassis obtained when you
configure the chassis with MAX
moduleSlot—1
channel—The same channel number from step 7