Chapter 4    Preparing LogTags for use 
37 
 
 
Enable pre-start logging  Pre-start logging is a feature that protects the recording in the 
event of a failure to actually press the START MARK button to 
initiate the recording. It is an insurance policy against the loss of 
critical data. This is almost always checked for transit monitoring 
applications. This option is further explained in the section "
Pre 
start data collection
". 
Record readings 
continuously, overwrite 
oldest when memory 
full 
This is what is called a “data wrap” in which the recorder just keeps 
writing data to memory. It is a “moving window” of data. For 
example, this is useful for fixed base MKT monitoring with the 
interval set to one year, so that the current “yearly MKT” can 
always be in the memory of the LogTag. If this option is selected, 
the user will be “opted out” of the next two options. This option is 
further explained in the section "
Continuous operation
". 
Record readings so 
that: 
If this is selected, the logger stops logging when the number of 
data points has been gathered and stored in memory that spans at 
least the number of days specified below. When this option is 
selected, the following two options will determine the number of 
data points over the span of time that will be recorded. 
Readings recorded will 
span at least X days 
If you know how long you want to record, the Configuration screen 
will automatically set the number of readings to be collected at a 
particular interval. This setting will only have an effect if the above 
option "Record readings so that:" is selected. 
Number of readings to 
record X maximum is Y 
This data box will display the calculated number of readings if you 
select a time span and an interval. Alternatively, you can select the 
number of readings desired, and the time span will be altered 
based on the interval. If this is selected, the logger will stop logging 
when the number of readings has been gathered and stored in 
memory. "Y" represents the maximum number of readings the 
LogTag can store, however with a selected interval and a selected 
time span, you may not need to use all of the available data space. 
This setting will only have an effect if the above option "Record 
readings so that:" is selected. 
Record a reading every 
X  
This determines how frequently the LogTag records the 
environmental humidity and/or temperature into its internal 
memory. 
Begin recording after a 
delay of X 
Determines how long after the START MARK button on the LogTag 
has been pressed that it will actually begin to record the 
environment humidity and/or temperature. This feature is useful for 
example, when the LogTag could falsely trigger an alert because 
the environmental conditions it is being started in is quite different 
to the environmental conditions to be monitored. If pre-start 
logging is enabled, the LogTag will continue to record pre-start 
readings during this delay. 
Enable the OK (Green) 
indicator 
This determines whether or not the green light should function 
while the LogTag is recording after being started. 
Humidity / Temperature  These tab controls will allow the humidity and temperature alert 
conditions to be configured independently. These tab controls will 
only appear when configuring a LogTag that has a humidity and a 
temperature sensor. If the LogTag has only one sensor then the 
following alert conditions will apply to that sensor.