Fluke 724 ユーザーズマニュアル

ページ / 56
724
Users Manual
30
Setting 0 % and 100 % Output
Parameters
For output parameters (volts, ohms, TC potentials or RTD
resistances), you must set the 0 % and 100 % points
before you can use the step and ramp functions. Proceed
as follows:
1.  If necessary, press 
M
 for SOURCE mode.
2.  Select the TC source function and use the arrow keys
to enter the value. Our example is thermocouple
source using 100 
°
C and 300 
°
C values for source.
3. Enter 
100 
°
C and press and hold 
J
 to store the
value.
4.  Enter in 300 
°
C and press and hold 
G
 to store the
value.
You can now use this setting for the following:
 
Manually stepping an output with 25 % increments.
 
Jump between the 0 and 100 % span points by
momentarily pushing 
J
 or 
G
.
Stepping and Ramping the Output
Two features are available for adjusting the value of
source functions.
 
Stepping the output manually with the 
H
 and 
I
keys, or in automatic mode.
 
Ramping the output.
Stepping and ramping apply to all functions.
Manually Stepping the Output
To manually step the output you can do the following:
 Use 
H
 or 
I
 to step the output up or down in
25 % steps.
 
Touch momentarily either 
J
 to go to 0 %, or 
G
to go to 100 %.