34
Testing an Output Device
Use the source functions to test and calibrate actuators,
recording, and indicating devices. Proceed as follows:
1.  Connect the test leads to the instrument under test
as shown in Figure 16.
2. Press 
l
 for dc voltage, or 
U
 for resistance
(lower display).
3.  If necessary, press 
M
 for SOURCE mode.
TC
RTD
100%
25%
25%
RECALL
MEAS
SOURCE
STORE
SETUP
0%
TEMPERATURE  CALIBRATOR
724
Black
Red
0 to 1 V dc 
Input
V
LOOP
mA
V
˚
F
˚
C
zi13f.eps
Figure 16. Calibrating a Chart Recorder