Temperature Calibrator
Specifications
39
Specifications
Specifications are based on a one year calibration cycle
and apply from +18 
°
C to +28 
°
C unless stated otherwise.
All specifications assume a 5 minute warmup period.
DC Voltage Measurement
Range
Resolution
Accuracy,
(% of Reading + Counts)
30 V
(upper
display)
0.001 V
0.02 % + 2
20 V
(lower
display)
0.001 V
0.02 % + 2
90 mV
0.01 mV
0.02 % + 2
Temperature coefficient -10 
°
C to 18 
°
C, +28 
°
C to
55 
°
C: 
±
0.005 % of range per 
°
C
DC Voltage Source
Range
Resolution
Accuracy,
(% of Reading + Counts)
100 mV
0.01 mV
0.02 % + 2
10 V
0.001 V
0.02 % + 2
Temperature coefficient -10 
°
C to 18 
°
C, +28 
°
C to
55 
°
C: 
±
0.005 % of range per 
°
C
Maximum load: 1 mA
DC mA Measurement
Range
Resolution
Accuracy,
(% of Reading + Counts)
24 mA
0.001 mA
0.02 % + 2
Temperature coefficient -10 
°
C to 18 
°
C, +28 
°
C to
55 
°
C
±
0.005 % of range per 
°
C
Drive capability: 1000 
Ω
 at 20 mA