Emerson 1056 ユーザーズマニュアル
45
MODEL FCL-1056
SECTION 6.0
CALIBRATION
6.4.4 Procedure-Manual Calibration
1. Obtain two buffer solutions. Ideally, the buffer pH values should bracket the range of pH values to be measured.
1. Obtain two buffer solutions. Ideally, the buffer pH values should bracket the range of pH values to be measured.
2. Remove the sensor from the flow cell. If the process and buffer temperatures are appreciably different, place
the sensor in a container of tap water at the buffer temperature. Do not start the calibration until the sensor has
reached the buffer temperature.
3.
Press MENU. The main menu screen appears. The cursor will be on
Calibrate. Press ENTER.
Calibrate. Press ENTER.
4.
Choose the sensor you wish to calibrate. Sensor 1 is the chlorine sensor.
Sensor 2 is the pH sensor.
Sensor 2 is the pH sensor.
5.
Choose pH.
6.
Choose Buffer Cal
7.
Choose Manual.
8.
Choose Buffer 1.
9.
Rinse the sensor with water and place it in the first buffer. Be sure the
glass bulb and reference junction are completely submerged. Swirl the
sensor.
glass bulb and reference junction are completely submerged. Swirl the
sensor.
Calibrate?
Sensor 1
Sensor 2
Output 1
Output 2
Sensor 2
Output 1
Output 2
S2 Calibration
pH
ORP
Redox
Temperature
ORP
Redox
Temperature
pH
S2 pH Cal
Zero Cal
Standardize
Slope: 56.19 mV/pH
Offset: 2 mV
Standardize
Slope: 56.19 mV/pH
Offset: 2 mV
Buffer Cal
S2 pH Buffer Cal
Auto
Manual
Manual
S2 pH Manual Cal
Buffer 1
Buffer 2
Buffer 2
Buffer 1
Sensor 2