Fisher & Paykel MR850 ユーザーズマニュアル

ページ / 62
 
MR850 Technical Manual   *   Revision J   * Issued November 2005   *   Ref.  185041340 (185041713 internal)
 
61
 
History Change for model GJU 
PCB 
Version 
Software 
Version 
Introduction 
Serial Number 
Comments 
4.40 
19 September 1998  9885xxx00000 
First production release. 
4.44 
22 October 1998 
 
Software upgrade. 
5.12 
12 April 1999 
9985xxx00053 
Release of new PCB to 
accommodate change in 
temperature probe circuit 
redesign. 
5.13 
5 May 1999 
 
Software upgrade. 
5.13 
14 January 2000 
2000-85xxx00028 
Release of new PCB to 
improve manufacture. 
5.23 
3 April 2000 
2000-85xxx01661 
Software upgrade for low flow 
control stability. 
5.33 
25 July 2000 
2000-85xxx02806 
Software upgrade, allowing 
Neonatal volume ventilation 
capability.  Manual 
Temperature Compensation 
(TC) implemented. 
5.34 
11 October 2000 
2000-85xxx04212 
Software upgrade to improve 
EMI immunity. 
5.45 
19 February 2001 
2001-85xxx00427 Software 
upgrade. 
6.00 
15 November 2001 
Only released as 
software upgrade 
kits. 
Software upgrade.  
Introduction of the automatic 
and manual Humidity 
Compensation (HC) mode.  
Added time delay to the OFF 
button.  Enable low 
temperature alarm in stand-by. 
7.00 
2 April 2002 
2002-85xxx00976 
Software upgrade. 
Improve HC speed in Non-
invasive mode. 
F&P Co-axial circuit 
recognition and control. 
7.20 
18 March 2004 
2004-85GJU002592 
Software upgrade. 
7.21 
9 June 2004 
2004-85GJU006775 
Software upgrade. 
PTS access through serial 
command. 
Humidifier model protection. 
Software release Japan only. 
7.21 
9 July 2004 
2004-85GJU008765 
Release of new PCB capable 
of selective soldering.