Input and Output Calibration for Recorder with Display 
12/03 
DR4300 Circular Chart Recorder Product Manual 
149 
5.4.4  RTD (Resistance Temperature Detector) Inputs 
 
Table 5-7  Set Up Wiring Procedure for Calibrating RTD Inputs 
Step Action 
Connect the copper wire to the calibration source (see Table 5-3). 
Connect the other end of the copper wire to the TB2 terminals on the printed circuit assembly 
for the input channel being calibrated.  
 
Main PCA 
Pen 2
Main PCA 
Pen 1
7
TB2
TB1
7
TB2
TB1
TB2
+
R
-
+
Decade box
Copper
wires
24093
 
Figure 5-4  Calibration Set Up Diagram for RTD Inputs