Appendix G: Product Verification
Refer to Table 4-1 for a list of the equipment required to verify the perfor-
mance of your OA 5000 Series instrument.
Table 4-1: Equipment Required for Performance Verification
Description
Minimum Use Specifications
Recommendation
850 nm
Laser source
CW, 850 ±20 nm, actual wavelength
known within ±1 nm; >500
W into
50/125
m fiber; <0.02 dB/minute
stability
1
Fotec S785
1310 nm
Laser source
CW, 1310 ±20 nm, actual
wavelength known within ±1 nm;
>1 mW into SM fiber; <0.02 dB/
minute stability
1
Advantest
Q81211/Q8221
1550 nm
Laser source
CW, 1550 ±20 nm, actual
wavelength known within ±1 nm;
>500
W into SM fiber; <0.02
dB/minute stability
1
Advantest
Q81212/Q8221
(1310 and 1550
nm) Optical
power meter
with long
wavelength
sensor
±(0.025 dB + 0.3 pW) linearity, --- 90
to +3 dBm; noise (p --- p) < 0.5 pW;
accepts 9/125 to 100/140
m fiber
Agilent/HP
81532A/8153A
(850 nm)
Optical power
meter with short
wavelength
sensor
±(0.025 dB + 0.3 pW) linearity,
--- 90 to +3 dBm; accepts 9/125 to
100/140
m fiber
Agilent/HP
81530A/8153A
OA 5002,
Optical input fi-
ber, 1 each
50 m long, if available, or a fiber as
short as 1 m can be used;
9/125
m, FC/PC to FC/PC
2
OA 5002,
Optical output fi-
ber, 1 each
1 to 5 m long, 9/125
m, FC/PC to
FC/PC
2
Tektronix
174-1387-00
OA 5012,
Optical input
and output fiber
1 to 5 m long, 50/125
m, FC/PC to
FC/PC
2
Equipment List