Tektronix OA 5000 ユーザーズマニュアル

ページ / 134
Commands
Programming
3 ---24
FACTORY (No Query Form)
The FACTORY command resets the OA 5000 to its factory default settings
and purges stored settings.
NOTE
The FACTORY command can take 5 to 10 seconds to complete
depending on attenuation settings.
DESE, *ESE, HEADER, *PSC, *RST, *SRE, VERBOSE.
FACTORY
FACTORY
The FACTORY command does the following:
H Puts the OA 5000 in the Operation Complete Command Idle State.
H Puts the OA 5000 in the Operation Complete Query Idle State.
H Clears the Event Status Enable Register (equivalent to the command
*ESE 0
).
H Clears the Service Request Enable Register (equivalent to the command
*SRE 0
).
H Sets the Device Event Status Enable Register to all-enabled (equivalent
to the command DESE 255).
H Sets the Power-on status clear flag to TRUE (equivalent to the command
*PSC 1
).
H Sets the Response Header Enable State to TRUE (equivalent to the
command HEADER 1).
H Sets the Verbose Header State to TRUE (equivalent to the command
VERBOSE 1
).
H Sets the front panel as shown in Table 3-6.
Table 3-6: FACTORY Front Panel Settings
Front Panel Parameter
Setting
DISP:DB
ATT:DB
0
REF
0
STORE1
0
STORE2
0
Related Commands:
Syntax: