CC-100 Coolant Circulator Manual
Version 1.B
Cooling Capacity
Temperature rises approximately 0.03°C per watt (3.4 btu/hr). Based on a typical
flow rate of between 0.5 (0.47 liter) to 1 quart (0.95 liter) per minute and a thermal
load of 100 watts (341 btu/hr).
flow rate of between 0.5 (0.47 liter) to 1 quart (0.95 liter) per minute and a thermal
load of 100 watts (341 btu/hr).
When using a CC-100 to cool a PI detector, the coldest attainable temperature should
be reduced by no more than two degrees compared with using tap water at ambient
temperature.
be reduced by no more than two degrees compared with using tap water at ambient
temperature.
Operating Environment
Temperature
Minimum:
0°F (-18°C). Avoid freezing coolant.
Maximum:
120°F (49°C).
Humidity:
Avoid condensation
Clearance
Front:
6 inches (150 mm) minimum
Rear:
12 inches (300 mm) minimum
Side:
1 inch (20 mm) minimum
Power Requirements:
50/60 Hz
2 A @ 105V to 125V AC (standard)
2 A @ 90V to 110 V AC (factory installed option)
1 A @ 210 V to 250 V AC (factory installed option)
2 A @ 90V to 110 V AC (factory installed option)
1 A @ 210 V to 250 V AC (factory installed option)
Mechanical
Dimensions
D, W, H: 12½
318mm) × 15¾ (400mm) × 8¼ (210 mm)
Weight (without coolant)
16 lb (7.3 kg) without transformer, standard
19 lb (8.6 kg) with optional transformers for either 100V or 230V AC operation
19 lb (8.6 kg) with optional transformers for either 100V or 230V AC operation
Fluid Interface
Two ¼
female NPT tapped threads provided at the Inlet and Outlet on the rear
panel.
Two ¼
male NPT connectors to ¼ swage tube fittings with two 3 (76mm) by ¼
OD copper tubes provided standard.
Two ¼
plastic Swagelock tube plugs are provided standard.
Options
Tube accessory kit with eight Swagelock quick disconnect connectors and 20 feet
(6 m) of polyurethane tubing ¼
(6 m) of polyurethane tubing ¼
OD by 3/16 ID.