SECTION 1—GENERAL GUIDELINES
14
Make sure there is an audible click when mounting battery on the battery charger to 
confirm proper mounting. Otherwise, injury or damage may occur.
DO NOT exceed maximum weight limitation of the patient lift. The weight limitation for 
the RPL450‐2 (Low Base) is 450 lbs and the RPA600‐2 (Low Base) is 600 lbs.
ALWAYS keep hands and fingers clear of moving parts to avoid injury.
Using the Sling
Use an Invacare approved sling that is recommended by the individual’s doctor, nurse or 
medical assistant for the comfort and safety of the individual being lifted.
DO NOT use any kind of plastic back incontinence pad or seating cushion between 
patient and sling material that may cause the patient to slide out of the sling during 
transfer.
After each laundering (in accordance with instructions on the sling), inspect sling(s) for 
wear, tears, and loose stitching.
Bleached, torn, cut, frayed, or broken slings are unsafe and could result in injury. Discard 
immediately.
DO NOT alter slings.
Be sure to check the sling attachments each time the sling is removed and replaced, to 
ensure that it is properly attached before the patient is removed from a stationary object 
(bed, chair or commode).
If the patient is in a wheelchair, secure the wheel locks in place to prevent the chair from 
moving forwards or backwards.
When connecting slings equipped with color coded straps to the patient lift, the shortest 
of the straps MUST be at the back of patient for support. Using long section will leave little 
or no support for patientʹs back. The loops of the sling are color coded and can be used to 
place patient in various positions. The colors make it easy to connect both sides of the 
sling equally. Make sure that there is sufficient head support when lifting a patient.
Lifting the Patient
When using an adjustable base lift, the legs MUST be in the maximum Opened/Locked 
position before lifting the patient.
When the sling is elevated a few inches off the surface of the bed and before moving the 
patient, check again to make sure that the sling is properly connected to the hooks of the 
swivel bar. If any attachments are not properly in place, lower the patient back onto the 
stationary surface and correct this problem ‐ otherwise, injury or damage may occur.
Adjustments for safety and comfort should be made before moving the patient. Patientʹs 
arms should be inside of the straps.