Pride Mobility INFMANU3355 ユーザーズマニュアル

ページ / 16
14
Basic Operation Instructions
GC Controller
www.pridemobility.com
CARE AND MAINTENANCE
Refer to your power chair owner’s manual for proper cleaning and disposal
instructions.
WARRANTY
For two (2) years from the date of purchase, Pride will repair or replace at our
option to the original purchaser, free of charge, the controller or any of its components
found upon examination by an authorized representative of Pride to be defective
in material and/or workmanship.
Fault
Probable Cause(s)
Possible Solution
Code
1 Flash
The batteries need charging.
Check the battery connections. If the
There is a bad connection
battery connections are good, then
to the batteries.
try charging the batteries.
2 Flashes
There is a bad connection on
Check the left motor connections and
the left motor.
 wiring harnesses.
3 Flashes
The left motor has a short to
Contact your authorized Pride
a battery connection.
Provider.
4 Flashes
There is a bad connection on
Check the right motor connections
the right motor.
and wiring harnesses.
5 Flashes
The right motor has a short to Contact your authorized Pride
a battery connection.
Provider.
6 Flashes
The charge inhibit has been
Make sure that the battery charger
activated.
is not connected to the power base.
7 Flashes
There is a joystick fault.
Make sure the joystick is in the center
position and then power off and on
again.
8 Flashes
There is a control system fault. Check power module and joystick
module connections and wiring.
9 Flashes
The park brakes have a bad
Check the left and right motor
connection.
connections and wiring.
10 Flashes
An excessive voltage has
Check the battery connections.
been applied to the control
system.