Leica B523 ユーザーズマニュアル

ページ / 18
 
 
 
Objective 
Numerical 
Aperture 
(NA) 
Color 
ring 
Coverglass  (c.g.) 
thickness 
Position 
for 
auxiliary 
condensor lens 
N Plan 2.5x 
0.07 
Brown 
Any/without c.g.  
off 
N Plan 5x 
0.12 
Red 
Any/without c.g. 
off 
N Plan 10x 
0.25 
Yellow 
Any/without c.g. 
on 
N Plan 20x 
0.4 
Green 
0.17 mm  
on 
N Plan 40x 
0.65 
Blue 
0.17 mm  
on 
C Plan 63x 
0.75 
Black 
0.17 mm  
on 
Table 1. Objectives in Leica DM LB 
N Plan stands for planachromat and C Plan for achromat objectives. The  achromatic 
objectives  are  corrected  for  axial  chromatic  aberration  in  two  wavelengths  (blue  and 
red; about 486 and 656 nanometers, respectively). Furthermore, achromatic objectives 
are corrected for spherical aberration for the green color (546 nanometers). If focus is 
chosen in the green region of the spectrum, images will have a reddish-magenta halo. 
Achromatic objectives yield their best results with light passed through a green filter.  
Planachromat  is  a  flat-field  corrected  achromat  objective,  which  corrects  the  field 
curvature.  
NA  indicates  the  numerical  aperture  of  an  objective.  This  determines  the  resolving 
power  of  the  objective.  The  higher  the  NA,  the  better  the  resolution.  None  of  the 
objectives are to be used with immersion oil. 
For  wider  illumination  when  using  2.5x  and  5x  objectives,  move  the  auxiliary 
condensor lens out of the light path by swinging the rod connected to the lens (Figure 
4). 
 
 
Figure 4. Auxiliary condenser lens switch 
2.5
 
Focusing 
Focus  on  the  sample  by  turning  either  of  the  focusing  wheels  located  identically  on 
both sides of the microscope (Figure 5). Focusing raises or lowers the specimen stage; 
the  outer  wheel  is  for  coarse  focusing,  and  the  inner  one  for  fine  focus  adjustment.