Leica B523 ユーザーズマニュアル

ページ / 18
 
 
Instruction 
 
View in eyepiece 
4.  Center  the  diaphragm  in 
the  field  of  view  by  rotating 
the 
condenser 
adjusting 
screws.  In  other  words,  you 
adjust  the  illuminating  path 
so  that  it  is  properly  aligned 
along  the  optical  axis  of  the 
microscope. 
5.  Open  the  luminous-field 
diaphragm  by  turning  the 
adjustment  wheel  counter-
clockwise until the edges just 
disappear  from  the  field  of 
view.  This  step  ensures  that 
you only illuminate the field-
of-view. 
 
6.  Pull  out  one  ocular  and 
look  through  the  eyepiece 
tube.  Close  the  aperture 
diaphragm  by  turning  the 
wheel  in  the  condenser  until 
the  aperture  is  about  80%  of 
the  diameter  of  the  objective 
pupil.  Replace  the  eyepiece. 
Another way of adjusting the 
aperture is by looking at your 
specimen  and  closing  the 
diaphragm  until  resolution 
and  contrast  are  optimal. 
Diaphragm  too  open:  poor 
contrast. 
Diaphragm 
too 
closed: poor resolution. 
When  adjusting  Koehler  illumination  for  higher  magnification  objectives,  repeat  the 
procedure.  For  lower  magnifications  (5x  and  2.5x),  the  condenser  performance  is 
changed  by  the  removal  of  the  swing-out  lens  from  the  light  pathway,  and  the 
condenser  aperture  diaphragm  no  longer  controls  the  numerical  aperture.  Fully  open 
the  aperture  diaphrgam  and  control  the  contrast  by  opening/closing  the  field 
diaphragm.