Put the Analyzer into State and Timing Modes
In this exercise you will use both the timing analyzer mode and the state
analyzer mode to correlate the state and timing measurements.  Pod 1 will be
capturing the state measurement and Pod 3 will be capturing the timing
measurement.  All other pods will be unassigned.
1
Press the 
Config
 MENU key.
2
In the Analyzer 2 box, highlight the field to the right of 
Type
, and press
the 
Select
 key.  Highlight 
Timing
 in the pop-up menu, and press the
Select
 key.
3
Highlight the Pod A3/A4 field, and press the 
Select
 key.  Highlight
Machine 2
 in the pop-up menu, and press the 
Select
 key.
4
If you have any other pod pair fields other than A1/A2 and A3/A4,
select each pair and choose  
Unassigned
 in the pop-up menu.
Putting the Analyzer into State and Timing Modes
Result
Analyzer 1 is in state mode with Pod 1 assigned to it.  Analyzer 2 is in timing
mode with Pod 3 assigned to it.  The remaining pods are unassigned.
Analyzer 1 is in state
mode with the lower 8
bits of Pod A1 and the J
clock are showing
activity.
Analyzer 2 in timing
mode.  The lower 8 bits
of Pod A3 and the L clock
are showing activity.
If you are using an HP 1660 or
an HP 1670, set Pod A7/A8 to
"Unassigned". 
If you are using a 1660, 1661,
1670, or a 1671 set Pod A5/A6
to "Unassigned".
Mixed Mode - Correlation of State and Timing Data
Put the Analyzer into State and Timing Modes
5-5