5 Managing Patients
Care Groups
90
Telemetry Data Overview Screen Element
WARNING
Waves and numeric information on the telemetry screen element are delayed for several seconds. If you 
need realtime data, for example for defibrillation, always use the monitor ECG instead of telemetry.
Silencing Telemetry Alarms from the Bedside
When a telemetry device is paired with the monitor, an audible status change notification is issued for 
new telemetry alarms. The volume can be configured. 
If configured, alarms generated from a paired telemetry device can be silenced at the bedside.
To silence telemetry alarms from the bedside,
1
select the telemetry screen element 
2
select the 
Silence Bed 
pop-up key
Depending on your configuration, the 
Silence Bed
 key may silence both telemetry alarms and 
bedside alarms.
Suspending Telemetry Alarms
When you select 
Pause Alarms 
or 
Alarms Off
 at the monitor, the alarms are off or paused for 
the bedside measurements.
Refer to the Information Center Instructions for Use for the Suspend/Pause alarms behavior of the 
Information Center and telemetry device. 
Using Standby
When you select Standby mode at the monitor, the bedside goes into Standby mode. 
Refer to the Information Center Instructions for Use for the Standby behavior of the Information 
Center and telemetry device. 
Telemetry 