C.O./CCO Warning Messages
13 Monitoring Cardiac Output
173
C.O./CCO Warning Messages
Warning messages contain important information about the C.O. measurement. 
C.O./CCO Safety Information
WARNING
Catheter constant:  Make sure that the arterial catheter constant for the measurement is appropriate 
to the catheter used.
Computation Constant:  Make sure that the computation constant for the measurement is 
appropriate to the injectate volume, injectate temperature and catheter type used.
IABP:  Do not perform transpulmonary thermodilution measurements on patients undergoing IABP 
treatment.
CCO accuracy: Accuracy of the CCO measurement and all the derived values may be influenced by 
patients with valve diseases or artificial valves.
C.O. and MRI: Do not use the Cardiac Output Interface Cable in Magnetic Resonance Imaging 
(MRI) Applications.
Aortic graft patients:  Do not use an arterial catheter in the arteria femoralis when it is contra 
indicated, for example, with patients who have an aortic graft.
C.O./CCO Warning Messages
Possible Causes
Next measurement erases 
older curve 
Six curves are stored, this is the maximum possible. If another 
measurement is stored, the oldest thermodilution curve will be erased.
Previous C.O. Setup Data 
replaced
A C.O. module or measurement server extension has been plugged in 
with different C.O. setup data from the previous data. The new C.O. 
setup data is read from the new C.O. device, and replaces the current 
data. The message disappears when the 
Start C.O.
 pop-up key is 
pressed. 
Verify the C.O. Setup 
Data
A new transpulmonary thermodilution catheter has been connected to 
the C.O. Interface Cable.
Check arterial pressure, 
CCO cal currently not 
possible
Poor or invalid pressure signal, for example if pressure was not zeroed.
Verify the Computation 
Constant
A new catheter has been plugged in, or the computation constant has 
been changed and 
Start C.O. 
has not been selected.
Previous Comp. Constant 
replaced
A new C.O. module or measurement server extension with a different 
computation constant from the current one has been connected. The 
new computation constant is read from the new C.O. device, and 
replaces the current one. The message disappears when 
Start 
C.O.
is selected.