Philips MP60/70/80/90 사용자 설명서
Using the EEG Impedance/Montage Window
18 Monitoring EEG
209
Choosing an EEG Electrode Montage
1
To activate one of the five pre-configured electrode montages, select the arrow beside the label in
the
the
EEG Impedance/Montage
window and choose a montage from the list.
2
Attach the electrodes as illustrated in the
EEG Impedance/Montage
window.
The five default electrode montage configurations can be modified and renamed in Configuration
Mode.
Mode.
The electrode locations are labeled according to the international 10-20 electrode placement system.
Changing the Impedance Limit
The impedance limit can be set for all electrodes simultaneously in the
Setup EEG
menu, or in the
EEG Impedance/Montage
window using the pop-up keys. If the limit is exceeded during
monitoring, an INOP will appear and the graphic impedance indicator will change.
To change the impedance limit, either
♦
use the pop-up keys that appear with the
EEG Impedance /Montage
window, or
♦
in the
Setup EEG
menu, select
Impedance Limit
to call up a list of selections between 1 and
30 kOhm, then select the required limit from this list.
Montage Name
EEG1+
EEG1- Label1
EEG2+ EEG2-
Label2
Mont.A
Fp1
T3
Fp1-T3
Fp2
T4
Fp2-T4
Mont.B
O1
T3
O1-T3
O2
T4
O2-T4
Mont.C
F3
C3
F3-C3
F4
C4
F4-C4
Mont.D
C3
P3
C3-P3
C4
P4
C4-P4
Mont.E
Fp1
T5
Fp1-T5
Fp2
T6
Fp2-T6