34 Installation and Specifications
Safety and Performance Tests
368
Electrosurgery Interference/Defibrillation/Electrostatic Discharge
The equipment returns to the previous operating mode within 10 seconds without loss of any stored 
data. Measurement accuracy may be temporarily decreased while performing electro-surgery or 
defibrillation. This does not affect patient or equipment safety. Do not expose the equipment to x-ray 
or strong magnetic fields (MRI).
Fast Transients/Bursts
The equipment will return to the previous operating mode within 10 seconds without loss of any 
stored data (M1032 within 30 seconds). If any user interaction is required, the monitor indicates with 
a technical alarm (INOP).
Restart time
After power interruption, an ECG wave will be shown on the display after 30 seconds maximum.
Frequency of transmitter
150 kHz to 80 MHz
150 kHz to 800 MHz
800 MHz to 2,5 GHz
Equation
for respiration and BIS:
for respiration and BIS:
for respiration and BIS:
Rated max. output power 
of transmitter (W)
Separation distance (m)
Separation distance (m)
Separation distance (m)
0.01
0.1 (0.4)
0.1 (0.4)
0.2 (0.7)
0.1
0.4 (1.1)
0.4 (1.1)
0.7 (2.2)
1
1.3 (3.5)
1.3 (3.5)
2.3 (7.0)
10
3.8 (11.1)
3.8 (11.1)
7.3 (22.1)
100
12.0 (35.0)
12.0 (35.0)
23.0 (70.0)
d
1.2 P
=
d
3.5 p
=
d
1.2 P
=
d
3.5 P
=
d
2.3 P
=
d
7.0 P
=