Emerson 2200S 사용자 설명서
![Emerson](https://files.manualsbrain.com/attachments/1ef3aaad462e024fde6cecf537a4c6c781334d44/common/fit/150/50/6dab323528eefa0342baa74a10c15551c3b956250cf9d62f2117dcdf42a7/brand_logo.jpeg)
112
Micro Motion
®
Model 2200S Transmitters
Troubleshooting
Inaccurate density reading
Problem with process fluid
• Use standard procedures to check
quality of process fluid.
Bad density calibration factors
• Verify characterization. See
Wiring problem
• Check the sensor circuitry. See
Bad flowmeter grounding
• See Section 10.14.2.
Slug flow
• See Section 10.17.
Sensor cross-talk
• Check environment for sensor with
similar (±0.5 Hz) tube frequency.
Plugged flow tube
• Check drive gain and tube frequency.
Purge the flow tubes.
Incorrect sensor orientation
• Sensor orientation must be
appropriate to process fluid. See the
installation manual for your sensor.
installation manual for your sensor.
RTD failure
• Check for alarm conditions and follow
troubleshooting procedure for
indicated alarm.
indicated alarm.
Physical characteristics of sensor have
changed
changed
• Check for corrosion, erosion, or tube
damage.
Temperature reading significantly
different from process temperature
different from process temperature
RTD failure
• Check for alarm conditions and follow
troubleshooting procedure for
indicated alarm.
indicated alarm.
Temperature reading slightly different
from process temperature
from process temperature
Sensor leaking heat
• Insulate the sensor.
Unusually high density reading
Plugged flow tube
• Check drive gain and tube frequency.
Purge the flow tubes.
Incorrect K2 value
• Verify characterization. See
Unusually low density reading
Slug flow
• See Section 10.17.
Incorrect K2 value
• Verify characterization. See
Unusually high tube frequency
Sensor erosion
• Contact Micro Motion.
Unusually low tube frequency
Plugged flow tube, corrosion, or erosion
• Purge the flow tubes.
Unusually low pickoff voltages
Several possible causes
• See Section 10.23.3
.
Unusually high drive gain
Several possible causes
• See Section 10.23.2.
Table 10-4
Process variables problems and remedies continued
Symptom
Cause
Suggested remedy