Invacare IRC5LXAW 사용자 설명서
SECTION 5—OPERATING INSTRUCTIONS
26
GREEN light (O
2
) ‐ Normal Operation.
YELLOW light (
) ‐ Call supplier IMMEDIATELY. You may
continue to use the concentrator unless instructed otherwise
by your supplier. Be certain that backup oxygen is nearby.
RED light ( )‐ Total unit shutdown. Switch IMMEDIATELY
to backup oxygen supply. Call supplier IMMEDIATELY
(FIGURE 5.7).
GREEN light ‐ with YELLOW light flashing ‐ Call supplier
IMMEDIATELY. Oxygen sensor malfunctioning; you may
continue to use the concentrator (FIGURE 5.7).
by your supplier. Be certain that backup oxygen is nearby.
RED light ( )‐ Total unit shutdown. Switch IMMEDIATELY
to backup oxygen supply. Call supplier IMMEDIATELY
(FIGURE 5.7).
GREEN light ‐ with YELLOW light flashing ‐ Call supplier
IMMEDIATELY. Oxygen sensor malfunctioning; you may
continue to use the concentrator (FIGURE 5.7).
Indicators (LX Model)
NOTE: For this procedure, refer to FIGURE 5.8.
Explanation of LX models Indicator Lights
RED light ( ) ‐ Total Unit Shut‐Down. Switch
IMMEDIATELY to a back‐up oxygen supply. Call supplier
IMMEDIATELY.
GREEN light (I/O) ‐ On/Off. System okay.
IMMEDIATELY to a back‐up oxygen supply. Call supplier
IMMEDIATELY.
GREEN light (I/O) ‐ On/Off. System okay.
O
2
Auto Shut Down
Use Backup
Below
Normal
Normal
RED
Call Supplier
YELLOW
GREEN
OXYGEN PURITY