Invacare 1148114 사용자 설명서

다운로드
페이지 24
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
13
2. Ensure toggle valve is open. Open the toggle valve by 
rotating counterclockwise.
3. Examine the gauge on the connected conserver or 
regulator. Perform one of the following:
• If the cylinder pressure is less than 1500 psig, fill the 
cylinder. Connect the cylinder to the compressor. Refer 
to Filling the Cylinder on page 13.
• If the cylinder pressure is greater than 1500 psig., DO 
NOT attempt to fill this cylinder. It may not fill. Use 
cylinder as normal.
Filling the Cylinder
NOTE: For this procedure, refer to FIGURE 4.2 thru FIGURE 4.7.
1. Check the cylinder pressure before attempting to fill 
2. Close the toggle valve by rotating clockwise (FIGURE 4.2).
3. Remove the conserver or regulator according to the device