ResMed Stellar 100 사용자 설명서
17
Setup menu
En
glis
h
Setup menu: Configuration Menu
Parameter
Description
Learn Circuit
The device is calibrated according to your air tubing system.
1. If in use, turn off the oxygen flow.
2. Select the mask type.
3. Set up the air circuit including accessories and patient interface.
1. If in use, turn off the oxygen flow.
2. Select the mask type.
3. Set up the air circuit including accessories and patient interface.
Note: For invasive use, do not connect a catheter mount,
tracheostomy tube or HMEF, see “Setting up for invasive use”
on page 5).
tracheostomy tube or HMEF, see “Setting up for invasive use”
on page 5).
4. Leave the air circuit unobstructed and open to the air.
5. Press
5. Press
to start the Learn Circuit.
6. Wait for the device to complete its automated tests (<30 sec).
The results are displayed when complete. If the circuit
configuration has been successfully learnt,
configuration has been successfully learnt,
displays. If
unsuccessful,
displays (see “Troubleshooting” on page 22).
FiO
2
Sensor
Calibration
The device starts the calibration of the FiO
2
monitoring sensor to
measure the oxygen concentration of the breathable air.
1. Press
1. Press
to start the FiO
2
sensor calibration.
2. Wait for the device to complete its calibration.
The results are displayed when complete.
Note: Turn off the oxygen flow.
Ramp Time
If Max Ramp Time has been set by the clinician, you may select any
value up to this time.
Options: 0 min-Max Ramp Time (max 45 minutes, 5 minutes
increments)
value up to this time.
Options: 0 min-Max Ramp Time (max 45 minutes, 5 minutes
increments)
Parameter
Description
Language
Sets the display language.
Options: Depending on regional configuration
Options: Depending on regional configuration
Brightness
Sets the LCD backlight brightness.
Options: 20-100%, 10% increments
Options: 20-100%, 10% increments
Backlight
Enables the LCD and keypad backlight.
If the AUTO setting is selected the backlight turns dark after five
minutes without any actions and turns on again if any button is
pressed again or an alarm occurs.
Options: On, Auto
If the AUTO setting is selected the backlight turns dark after five
minutes without any actions and turns on again if any button is
pressed again or an alarm occurs.
Options: On, Auto
Time format
Sets the time format.
Options: 24 hrs,12 hrs
Options: 24 hrs,12 hrs
Date format
Sets the date format.
Options: dd/mm/yyyy, mm/dd/yyyy
Options: dd/mm/yyyy, mm/dd/yyyy
To return to the Options screen press the Return button