104 Programming the VT1422A for Data Acquisition and Control  
Executing the Programming Model
This section shows the sequence of programming steps that should be used 
for the VT1422A. Within each step, most of the available choices are shown 
using command sequence examples, with further details available in the 
Command Reference Chapter 7.
 IMPORTANT!
It is very important that, while developing an application, the 
SYSTem:ERRor? query be executed after each programming command. 
This is the only way to determine if a programming error has occurred. 
SYST:ERR? returns an error number and description (or +0, "No Error").
Power-on and *RST
Default Settings
Some of the programming operations that follow may already be set after 
Power-on or after a *RST command. Where these default settings coincide 
with the configuration settings required, there is no need to execute a 
command to set them. These are the default settings:
No algorithms defined
No channels defined in channel lists
Programmable SCPs configured to their Power-on defaults 
(see individual SCP User’s Manuals)
All analog input channels linked to EU conversion for voltage
All analog output channels ready to take values from an algorithm
All digital I/O channels set to input static digital state
ARM:SOURce
IMMediate
SAMPle:TIMer
40E-6 (40 µs)
TRIGger:SOURce
TIMer
TRIGger:COUNt
(Note that this default was chosen to make testing data acquisition 
scan list easier. For algorithm operation, changing the count to 
INFinite is probably desired.)
TRIGer:TIMer
.010
(10 ms)
FORMat
ASC,7
(ASCII)
SENSe:DATA:FIFO:MODE
BLOCking
Figure 4-4 provides a quick reference to the Programming model. Refer to 
this along with the “Programming Overview Diagram” to keep an overview 
of the VT1422A SCPI programming sequence. Again, when default settings 
match the settings desired, that configuration step can be skipped.