124 Programming the VT1422A for Data Acquisition and Control  
ROUTe:SEQuence:DEFine accepts both on-board channels from 
conventional SCPs as well as remote channels from Remote Signal 
Conditioning Units (RSCUs). For details about syntax see “Channel List 
(Standard Form)” startin
g on page 231 and “ROUTe:SEQuence:DEFine” 
on page 323.
Note
Certain analog input SCPs display higher than normal offset and noise 
figures if their channels are scanned just before channels on a Remote 
Signal Conditioning Unit. To avoid any such interaction, the scan list 
should be order in such a way that all remote channels (5-digit channel 
numbers) appear before any on-board channels (3-digit channel numbers).
Example Scan List
To set up a scan list to take measurements on all on-board channels of a 
conventional SCP in position 0 and all remote channels of four VT1529A/Bs 
connected to two VT1539A SCPs in SCP positions 1 and 2:
ROUT:SEQ:DEF (@100:107,10800:11731)
Controlling Scan List
Data Destination
Readings taken on channels specified by ROUT:SEQ:DEF by default go to 
both the FIFO buffer and the CVT. By using another form of the <ch_list> 
parameter this data destination can be controlled to be the CVT, the FIFO, 
or even neither (no reading stored). For more on controlling data destination, 
See “ROUTe:SEQuence:DEFine” on page 323.
Example Scan List with controlled data destination
To set up a scan list as above but send the remote channel readings only to 
the FIFO buffer:
ROUT:SEQ:DEF (@100:107,2(10800:11731))