VXi VT1529A/B 사용자 설명서

다운로드
페이지 529
Advanced Programming with the VT1529B   165
Field Wiring for
Excitation
Measurements
To take advantage of the on-the-fly excitation measurement feature means 
that at least one channel of each VT1529B will need to dedicated to the 
excitation voltage measurement. Any of the 32 channels can be used for this 
purpose. If the application requires more frequent measurements than once 
per scan, multiple channels can be dedicated and a separate excitation 
measurement channel for each strain measurement channel can even be 
used. The excitation measurement channel must be specified for each strain 
measurement channel using SENS:FUNC:STR:<bridge_type>:POST.
For the dedicated excitation measurement channel, simply leave the RJ-45 
ports for the channel unconnected. (Putting a short between the strain gage 
Sense + and Sense - inputs, however, may be desired. See “Settling Time 
Considerations” on page 178
 for a discussion of why this may be desired.) 
This works because there is an internal 10 k
Ω resistor between the Bridge 
Excitation input on the front of the VT1529B and the Excitation Sense + and 
- inputs for each channel in the 8-channel bank that each Bridge Excitation 
input provides power to. Thus, making an excitation voltage measurement 
on an unconnected channel input will measure the voltage supplied to the 
Bridge Excitation input.
Note
For systems with multiple VT1529Bs, the excitation measurements must be 
made on the same VT1529B as the strain gage channels that use that 
excitation voltage. This means that for a system with two VT1529Bs, the 
excitation voltage cannot be measured on only one VT1529B with that 
measurement being applied to the strain gage measurements on both 
VT1529Bs.
Note
The wiring for the strain gage channels is unchanged.
Strain Measurement
Command
Sequence
The steps for making strain measurements with the new VT1529B 
functionality may already be familiar. The unstrained voltage and gage 
factor for the strain channels need to be provided. The excitation channel for 
each of the strain channels must also specified as well as the channels to be 
scanned. One difference is that the excitation channel(s) must be included in 
the scan list. Figure 5-1 shows the sequence of VXIplug&play driver calls 
to configure and scan strain channels using the new functionality.