168 Advanced Programming with the VT1529B  
Alternate Method of
Computing Strain
As stated above, to use the additional features of the VT1529B, the EU 
conversions are normally done in the VXIplug&play driver, not by the 
on-board DSP. For channels programmed using the new capability, all 
measurement data in the FIFO, CVT, and algorithms is voltage, not strain. 
For full- and half-bridge strain channels, there is an alternate method of 
doing the EU conversions if having the strain engineering units in the 
algorithms is needed while still using the most recent value of the excitation 
voltage for more accurate data. This method involves making strain 
measurements using the DSP while also scanning the excitation voltage. 
Because the full- and half-bridge strain conversions are linear with respect 
to the excitation voltage, an algorithm can be created that uses the ratio of 
the initial and current excitation voltages to correct the DSP-computed value 
to get a more accurate value.
Note
This method cannot be used for quarter-bridge strain channels because the 
EU conversion for quarter bridge is non-linear with respect to the excitation 
voltage. If a quarter-bridge strain measurements is required in an algorithm, 
the full EU conversion must be done in the algorithm.
Here are the steps to use this method:
1.  Configure the channels as one would for the VT1529A. The following 
simplified command sequence assumes that the strain channels are 10000 
through 10002 and that the common excitation voltage for these channels is 
monitored on channel 10006. It also assumes that the excitation voltage 
supply is short-term stable, so that the supply variation is negligible between 
the time the first and last channels are measured.
MEAS:VOLT:EXC? (@10000:10002)
Measure excitation on strain 
channels - the FIFO data needs 
to read and saved for later
MEAS:VOLT:UNST? (@10000:10002)
Measure unstrained voltages
SENS:STR:GFAC 2.2,(@10000:10002)
Specify gage factor
SENS:FUNC:STR:FBEN (@10000:10002)
Setup for full-bridge strain on 
channels (DSP will do EU)
2.  To have only the corrected value appear in the FIFO and CVT, disable 
sending the uncorrected values to the FIFO and CVT using the ROUT:DEF 
command:
ROUT:DEF (@(0(10000:10002),3(10006)
Disable sending of uncorrected 
strain data to FIFO and CVT; 
enable sending to both for 
excitation data
Alternatively, simply omit the channels for the uncorrected values from the 
scan list – the VT1422A driver will automatically add them to the scan list 
because they appear in the algorithm and won’t send them to the FIFO or 
CVT.